Perkin-Elmer PHI 660 - AES - Status: Down
|
Scanning Auger Multiprobe:
is a common analytical technique used specifically in the compositional study of surfaces. TRAINING READING REQUIREMENTS: Chapter 5.3 "Encyclopedia of Materials Characterization” Publisher: Butterworth-Heinemann ISBN-10: 0750691689 ISBN-13: 978-0750691680 Copy available on request. |
Scanning Auger Multiprobe:
Typical Applications: Surface Analysis and Depth Profiling
Signal Detected: Auger electrons
Elements Detected: Li-U
Detection Limits: 0.1 – 1 at%
5 – 50 Angstroms
Imaging/Mapping: Yes
Lateral Resolution/Probe Size: 5,000 Angstroms
The AES uses a finely focused electron beam to determine the elementals present in the top 3-4 atomic layers on a solid surface. All elements except H and He can be detected but the sample must be electrically conducting or semi conducting.
The Perkin-Elmer PHI 660 scanning Auger microprobe uses a LaB6 filament to probe the surface of a solid sample generating Auger electrons originating from the top surface of a sample in an ultra high vacuum (<10E-9 torr). It is capable of producing elemental maps or lie scans with a lateral resolution of less than 1 um. The machine is equipped with an Ar ion gun to remove surface layers and produce a depth profile of the distribution of the elements in the Z direction. Also available is an in situ fracture stage to create a fresh surface under UHV conditions.
An active Gatorlink login is required to create a new reservation.